北京中教金源科技有限公司
高级会员第2年 参观人数:160651
  • 参考报价:电议
    型号:
    产地:美国
    在线咨询
  • 详细介绍:


    Reference Solar Cell and Meter

    Reference Solar Cell and Meter

    Reads calibrated sun irradiance and temperature

    Easily integrated with I-V Test Station

    The Reference Cell, which is an integral part of solar simulator calibration and solar cell I-V characterization, consists of a readout device and a 2 x 2 cm calibrated solar cell made of monocrystaline silicon. The cell is also equipped with a thermocouple assembled in accordance with IEC 60904-2. The Reference Cell comes with calibration data taken by NREL, on an accompanying certificate. It reads solar simulator irradiance in "sun" units; one sun is equal to 1000 W/m2 at 25 °C and Airmass 1.5 Global Reference. This tool's primary use is the testing of photovoltaic cells under standard conditions. The Readout Meter includes two BNC connectors for analog outputs for the sun irradiance and the temperature.

    Specifications

    General

    Dimensions: 5.9 W x 3.8 H x 7.0 D inches (151x 95x178 mm)

    Weight: 4 lb. (1.8 kg)

    Operating Temperature: 10 ºC - 40 ºC

    Operating Humidity: 0 – 90%RH Non-Condensing

    Irradiance Readout

    Range: 0 – 3.500 Sun

    Accuracy: ±0.1% @ 1.0000 Sun @ 23 ºC

    ±0.2% +2 Counts @ 0.1000 – 1.9500 Sun (Low Display Range) @ 23ºC

    ±0.2% +2 Counts @ 1.900 – 3.500 Sun (High Display Range)@ 23 ºC

    Resolution: 0.0001 Sun @ 0 – 1.9500 Sun (Low Display Range)

    0.001 Sun @ 1.900 – 3.500 Sun (High Display Range)

    Temperature Coefficient: ±150 ppm / ºC Max

    Settling Time: <1 sec. for <0.25% (= 6ô)

    Sampling Rate: 3 Readings / second

    Autoranging: Switches to higher display range above 1.950 Sun, lower range below 1.900 Sun

    Analog Outputs

    Sun Irradiance: 1 volts/sun

    Temperature: 10 mV / 1ºC

    Temperature Readout

    Range: 0 – 199.9 ºC

    Accuracy: ±0.5 ºC Typical, ±1.0 ºC Max @ 24 – 26 ºC

    ±0.7 ºC Typical, ±1.2 ºC Max @ 10 – 40 ºC

    Resolution: 0.01 ºC

    Temperature Coefficient: ±0.02 ºC / ºC

    Settling Time: <1 sec. for <0.25% (= 6ô)

    Sampling rate: 3 Readings / second