深圳凯米斯科技有限公司
高级会员第2年 参观人数:23198
  • 参考报价:电议
    型号:
    产地:美国
    在线咨询
  • 详细介绍:


    此款手腕式振荡器已改进用于满足现行美国药典(USP)中二甲基硅油消泡性能测试。

    技术参数:

    半径距离:13.3±0.4cm

    容器数量:2 x 2

    振荡频率:300Osc/min

    幅度:10o

    **负载:10 lbs

    外型尺寸20 x 19 x 18

    电源:220V 50Hz

    重量:68 lbs

    选配:

    附:二甲基硅油(Simethicone)消泡性能美国药典测试方法

    Defoaming Activity

    Foaming solution: 10 mg/mL of octoxynol 9 in water

    Sample solution: 4 mg/mL of Simethicone in tertiary butyl alcohol. Shake vigorously.[Note—Warm slightly, if necessary, to dissolve. ]

    Analysis[Note—For each test, use a clean, unused, 250-mL glass jar. ]

    Add, dropwise, 500 µL of the Sample solution to a clean, unused, cylindrical 250-mL glass jar, fitted with a 50-mm cap, containing 100 mL of the Foaming solution. Cap the jar, and clamp it in an upright position on a wrist-action shaker. Using a radius of 13.3±0.4 cm (measured from the center of the shaft to the center of the bottle), shake for 10 s through an arc of 10 at a frequency of 300 Osc/min; 30 strokes/min. Record the time required for the foam to collapse. The time, in s, for foam collapse is determined at the instant the first portion of foam-free liquid surface appears, measured from the end of the shaking period.

    Acceptance criteria: The defoaming activity time does not exceed 15 s.