    Multicomponent FTIR Gas Analyzer

    多组分傅里叶红外光谱分析仪

    Gasmet On-line Series includes industrial multicomponent gas analyzers for continuous monitoring applications. The Gasmet CX4000 incorporates a Fourier transform infrared spectrometer, a temperature controlled sample cell, and signal processing electronics. The analyzer is fully equipped for fixed installations and it offers versatility and high performance for all industrial users.

    Gasme在线测试系列包括工业多组分气体分析仪应用于连续在线监测领域。Gasmet CX4000 由傅里叶变换红外光谱仪、 样品温度控制单元和信号处理电子系统组成。该分析仪为所有工业用户提供多功能和高性能的成套装备。

    The Gasmet CX4000 is designed for continuous emission monitoring. It is an ideal tool to analyze trace concentrations of pollutants in wet, corrosive gas mixtures. The sample cell can be heated up to 180 °C. Sample cell absorption path length is selected according to the application.

    Gasmet CX400分析仪是专门为检测烟气的连续排放而设计的。它是来分析潮湿的腐蚀性气体混合物中痕量污染物浓度的理想工具。样品池可以被加热到180 ℃,吸收光谱样品单元可根据客户实际应用调整选择。

    The Gasmet CX4000 allows simple calibration using only single component calibration gases. The user can easily configure the analyzer for a new set of compounds.

    Gasmet CX4000分析仪允许使用只组分标准气体标定。使用者很容易配置新的测试气体组分。

    General parameters 常规参数

    Measuring principle:Fourier transform infrared, FTIR

    Performance:Simultaneous analysis of up to 50

    gas compounds

    Response time, T90:Typically < 120 s, depending on

    the gas flow and measurement

    Time

    Operating temperature:5 - 30°C, non-condensing

    air conditioning recommended

    Storage temperature:-20 - 60°C, non-condensing

    Power supply:100-115 or 230 V / 50 -60 Hz

    Power consumption:300 W max, continuous 150

    测量原理:FT-IR(傅里叶变换红外光谱分析)

    性能:可同时测量多达50种气体组分

    响应时间T90:< 120 s,取决于气体流速和测量时间

    工作温度:5 - 30°C,无冷凝,建议使用空调以保持恒温。

    贮存温度:-20 - 60°C,无冷凝

    电源:100-115v or 230 V / 50 -60 Hz

    功率:**功率为300W,连续工作功率为150W

    Spectrometer 分光仪

    Resolution:4 cm-1

    Scan frequency:10 scans / s

    Detector:Peltier cooled MCT

    Source:SiC, 1550 K

    Beamsplitter:ZnSe

    Window material:ZnSe

    Wave number range:500 - 4 200 cm-1

    分辨率:4 cm-1

    扫描频率:10 次/s

    检测器:Peltier 制冷 MCT

    红外光源:SiC, 1550 K

    分束器:ZnSe

    窗口材质:ZnSe

    波数范围:500 - 4200 cm-1

    Sample cell 样品室

    Structure:Multi-pass, fixed path length 5.0 m

    Standard material:100 % rhodium coated aluminium

    Mirrors:Fixed, protected gold coating

    Volume:0.4 liters

    Connectors:Inlet Swagelok 6 mm

    Outlet Swagelok 8 mm

    Gaskets:Viton® O-rings

    Temperature:180 °C, maximum

    Window material:BaF2

    结构:多册反射光程,固定路径长度为5.0m

    标准材质:100%铑镀铝

    反射镜:固定,黄金镀层保护

    容积:0.4L

    连接器:入口接头套管6mm,出口接头套管8mm;

    垫片:O型氟化橡胶

    温度:**180℃

    窗口材质:BaF2

    Measuring parameters 测量参数

    Zero point calibration:24 hours, calibration with nitrogen (5.0 or higher N2 recommended)

    Zero point drift:< 2 % of measuring range per zero point calibration interval

    Sensitivity drift:None

    Linearity deviation:< 2 % of measuring range

    Temperature drifts:< 2 % of measuring range per 10K temperature change

    Pressure influence:1 % change of measuring value for 1 % sample pressure change.Ambient pressure changes measured and compensated

    零点校准:24H,用纯N2校准(推荐用5.0或更高的N2)

    零点漂移:< 2 %(小于每个零点的校准区间量程< 2 %)

    灵敏度漂移:无

    线性偏差:< 2 %(测量量程的2 %);

    温度漂移:每10K的温度变化内漂移量<量程2%

    压力影响:样气压力变化1%,测量值变化1%,环境压力变化可被监测并得以补偿。

    Electrical connectors 电连接器

    Digital interface:9-pole D-connector for RS-232 Analyzer is connected to an external computer via RS-232C cable. The external computer controls Gasmet.

    Power connection:Standard plug CEE-22

    数字接口:RS-232 D型9孔,分析仪通过 RS-232C 连接到外部计算机,由外部计算机控制傅立叶分析仪。

    电源连接:CEE-22标准插头

    Gas inlet and outlet conditions 气体进出口条件

    Gas temperature:Non-condensing, the sample gas temperature should be the same as the sample cell temperature

    Flow rate:120 - 600 liters per hour

    Gas filtration:Filtration of particulates (2 μm) required

    Sample gas pressure:Ambient

    Sample pump:External, not included

    气体温度:无凝结,样气温度必须和样品室的温度相同

    流速:120-600L/h

    气体过滤器:过滤精度2 μm

    样气压力:大气压

    样气泵:外置,不包含

    Electronics 电气

    A/D converter:Dynamic range 95 dB

    Signal processor:32-bit floating point DSP 120 MFLOPS

    Computer:External, not included

    A/D转换器:动态范围:95 dB

    信号处理器:32 bit浮点 DSP 120 MFLOPS

    计算机:外置,不包含

    Analysis software (for external PC)

    Operating system:Windows 7 (32-bit)

    Analysis software:Calcmet for Windows

    分析软件(外部计算机)

    操作系统:Windows 7 (32位)

    分析软件:基于Windows的Calcmet

    Options 选项

    Pressure measurement:Inside sample cell

    External PC:

    Gasmet PC for control and analysis with optional alarm relay boards and analog inputs or outputs

    Sample cell gaskets:Kalrez®

    压力测量:在样品室内部

    外置计算机:Gasmet工业电脑配备警报继电器板和模拟信号输入输出接口,用于系统控制和信号分析。

    Enclosure 附件

    Sample cell material:Aluminium

    Dimensions (mm):482 * 196 * 450

    Weight:17 kg

    CE label:According to EMI guideline 89/336/EC

    样品室材质:铝

    尺寸(mm):482 * 196 * 450

    重量:17KG

    CE标签:据 EMI 89/336/ec 号指令