北京盛田嘉源科技有限公司
高级会员第2年 参观人数:58510
  • 参考报价:电议
    型号:
    产地:美国
    在线咨询
  • 详细介绍:


    仪器简介:

    General Specifications

    Code N Performance

    Code C Performance

    Accuracy (combined nonlinearity, hysteresis and Nonrepeatability, % of full scale)

    ±0.2

    ±0.15

    Rotational Effect on Zero (% of full scale)

    ±0.05

    ±0.025

    Temperature Effects:

    Zero (% of full scale/deg F.)

    ±0.007

    ±0.003

    Span (% of reading/deg F.)

    ±0.007

    ±0.003

    Compensated Range

    +75 to +150 deg F

    Minimum Usable Range

    +32 to +175 deg F

    Analog Output

    ±10 or ±5 Volts dc, user selectable

    Nominal Overrange (% of full scale)

    30

    Bessel Response Signal Filter Cutoffs

    0.1 to 200 Hz in eleven 1-2-5 steps, user selectable

    System Resolution (% of full scale)

    0.02% on ±10V & 0.04% on ±5V

    RS232 Communications Port

    Outputs Torque in engineering units. Inputs units of measure selection, scaling, signal filter selection and cal data.

    Supply Voltage

    10 to 15 volts dc @ 200 mA, nominal

    技术参数:

    扭矩测量范围:2.83Nm~1130Nm

    信号输出:+/-5V,+/-10V, RS232C

    其他信息请您咨询我们公司。

    010-82413547 82416475

    Email: sale@sent*********

    主要特点:

    These Compact Digital Torquemeters have high accuracy, low noise, high overload capacity, high overrange, inherent noise tolerance and a wide temperature range. Strain gage sensing and non-contact signal transfer make them well suited for control, laboratory and production use. They are offered in two accuracy grades, Code N standard performance and Code C enhanced performance. Installation can be either floating or foot mounted with the integral mounting base.

    Null, scaling and units of measure are stored in non-volatile memory. There are no noisy pots or switches. Ten common units of measure are supported without requiring recalibration. Eleven user selectable Bessel filters avoid delay distortion and overshoot errors. Included software interfaces Windows-based PC's and displays and plots real time data. Use it to select ±5V or ±10V analog outputs, filters, scaling, units of measure and/or to control measurements.