北京康林科技有限责任公司
高级会员第2年 参观人数:50266
  • 参考报价:电议
    型号:
    产地:其他国家
    在线咨询
  • 详细介绍:


    技术参数:

    PSD/3 Specifications:

    Power Supply:

    Voltage: 24 VDC (±5%)

    Current: 1 amp typical; 2 amp peak

    Syringes:

    Syringe Type: For 4.5 in module, Hamilton GASTIGHT® Half Height Instrument Syringes

    Sizes: 12.5 uL, 25 uL, 50 uL, 125 uL, 250 uL, 500 uL, 1.25 mL, 2.5 mL, 5 mL, 12.5 mL

    Fluid Contact: Glass, PTFE, CTFE

    Syringe Drive:

    Drive Type: Closed loop DC Servo Motor: self lubricating and wear compensating lead nut/screw (no lubrication)

    Travel/speed: 30mm: 0.5seconds to 60 minutes for full stroke. Dependent upon system configuration.

    Resolution: 30,000 increments full stroke on 30 mm stroke length

    Adjustable to the following counts:

    Standard/half 1,000 counts

    Standard/full 2,000 counts

    High/half 15,000 counts

    High/full 30,000 counts

    Inaccuracy: ±1.0% @ > 30% stroke; ±1.2% @ 5% stroke (syringe dependent)

    Imprecision +0.2% @ > 30% stroke; ±0.5% @ 5% stroke (syringe dependent)

    Valves:

    Valve Type: Hamilton HV plug valves; snap-in/out

    Port Size: .039" (1.0 mm), .059" (1.5 mm), or .070" (1.8mm)

    Positions: 2 position @ 90° typical; multiple position possible (15° increments)

    Fluid Contact: PTFE & CTFE

    Fittings: Hamilton 1/4" - 28 male tube fittings with flared FEP or PTFE tubing

    Valve Drive:

    Drive Type: Closed loop DC motor with encoder

    Speed: 390 ms @ 90°

    Environment

    Operating Temperature: 60°F to 95°F (15°C to 35°C)

    Operating Humidity: 20% to 90%, non-condensing

    Storage Temperature: -5°F to 158°F (-20°C to 70°C)

    Storage Humidity: 20% to 95%, non-condensing

    Communication

    Protocol: Hamilton Reno Protocol 1 or Hamilton Bonaduz DIN Protocol

    Interface: RS-232C

    Baud Rate: 9600

    Daisy Chain: 16 modules

    Software Features:

    Memory Storage: FLASH

    Stored Methods: 1

    TTL I/O: 1

    Auxiliary Input: 1 (e.g. hand probe)

    Step/Prime Modes: Yes

    Firmware Revision: Programmable FLASH (In-System)

    Syringe Program Variables: Position, Speed, Prime Speed (system default), Backlash Compensation (return steps), & On The Fly Speed Change

    Syringe Program Variables: Delays Overload/Zero Sensitivity TBD through testing

    Valve Program Variables: Position and Speed

    PSD/3 Dimensions: Click here for a larger image.